5 kb (e5.1) technical data – BINDER KB 23 User Manual

Page 88

Advertising
background image

KB (E3.1 + E5.1) 11/2014

page 88/100

20.5 KB (E5.1) technical data

Unit size

240

400

720

Exterior dimensions
Width

mm / inch 925 / 36.42 925 / 36.42 1249 / 49.17

Height (incl. castors)

mm / inch 1460 / 57.48 1945 / 76.57 1924 / 75.75

Depth

mm / inch 800 / 31.50 800 / 31.50 887 / 34.92

including door handle, I-panel, connection

mm / inch 850 / 33.46 850 / 33.46 939 / 36.97

Wall clearance rear (minimum) (spacer)

mm / inch 100 / 3.94 100 / 3.94 100 / 3.94

Wall clearance side (minimum)

mm / inch 100 / 3.94 100 / 3.94 100 / 3.94

Steam space volume

l / cu.ft.

348 / 12.3 564 / 19.3 918 / 32.4

Quantity of doors

1

1

2

Quantity of inner glass doors

1

1

2

Interior dimensions
Width

mm / inch 650 / 25.59 650 / 25.59 970 / 38.19

Height

mm / inch 785 / 30.90 1270 / 50.00 1250 / 49.21

Depth

mm / inch 485 / 19.09 485 / 19.09 576 / 22.68

Interior volume

l / cu.ft.

247 / 8.7

400 / 14.1 698 / 24.6

Quantity of racks standard / max.

2/9

2/15

2/15

Load per rack

Kg / lbs

30 / 66

30 / 66

45 / 99

Permitted total load

Kg / lbs

100 / 220

120 / 265

150 / 331

Weight (empty)

Kg / lbs

170 / 375

220 / 485

309 / 681

Temperature data
Temperature range 1)

°C / °F

-5 to +100 /

23 to 212

-5 to +100 /

23 to 212

-5 to +100 /

23 to 212

Temperature fluctuation

max.

± K

0.1

0.1

0.1

Temperature uniformity

(variation)

max.

± K

0.5

0.6

0.5

at 4 °C / 39.2°F

± K

0.2

0.4

0.2

at 25 °C / 77°F

± K

0.2

0.2

0.2

at 37 °C / 98.6°F

± K

0.2

0.3

0.2

Max. heat compensation up to 40 °C / 104°F

W

300

500

600

Recovery time after doors

were open for 30 sec. 2)

at 4 °C / 39.2°F minutes

16

14

12

at 25 °C / 77°F minutes

1

1

1

at 37 °C / 98.6°F minutes

1

3

2

Electrical data
IP System of protection acc. to EN 60529

20

20

20

Nominal voltage (+/-10%)

V

200-240

1N~

200-240

1N~

200-240

1N~

Power frequency

Hz

50

50/60

50/60

Power plug

shock proof

plug

shock proof

plug

shock proof

plug

Nominal power

kW

1.20

1.40

2.30

Energy consumption 3)

at 4 °C / 39.2°F

Wh/h

245

385

435

at 25 °C / 77°F

Wh/h

225

365

420

at 37 °C / 98.6°F

Wh/h

260

420

510

Overvoltage category acc. to IEC 61010-1

II

II

II

Pollution degree acc. to IEC 61010-1

2

2

2

Unit fuse 5x20 mm 230V / middle-time-lag M

Amp

16 external 16 external 16 external

Noise level (approx.)

dB (A)

52

53

53

Advertising
This manual is related to the following products: