Performance in case of failures, 1 performance after a power failure, 2 alarm messages – BINDER KB 23 User Manual

Page 65: Temperature safety devices, 1 over temperature protective device (class 1)

Advertising
background image

KB (E3.1 + E5.1) 11/2014

page 65/100

13. Performance in case of failures

13.1 Performance after a power failure

Power failure during fixed-value operation (Normal Display):
The entered parameters remain saved. After power returns, operation continues with the set parameters.
Power failure during program operation:
After power returns, program course continues with the set-points that have been reached previously

during program operation.

13.2 Alarm messages

Alarm messages, e g. “TEMP. LIMIT” when exceeding the tolerance limit, are shown in Display 2 only in

Normal Display.
A buzzer can be activated / deactivated in the user level (chap. 10). It can be reset by pressing the

EXIT

button. The alarm text shown in Normal Display goes off only if the cause of the alarm does not exist any

longer.

14. Temperature safety devices

14.1 Over temperature protective device (class 1)

The cooling incubator is equipped with an internal temperature safety device class 1 acc. to DIN 12880. It

serves to protect the unit and prevents dangerous conditions caused by major defects.
If a temperature of about 110 °C / 230°F is reached, the over temperature protective device permanently

turns off the unit. The user cannot restart the device again. The protective cut-off device is located

internally. Only a service specialist can replace it. Therefore, please contact an authorized service provider

or BINDER Service.

14.2 Safety controller (temperature safety device class 3.1)

The cooling incubator is equipped with an electronic safety controller (over temperature safety device

class 3.1 acc. to DIN 12880). This second, electrically independent temperature controller takes over at a

selectable set point in case of a faulty condition. It serves to protect the charging material against

extremely high temperatures.
The safety controller serves to protect the cooling incubator, its environment and the material under

treatment from excess temperatures. In case of error the temperature of the inner chamber will be limited

to the set value. Please observe the guideline BGI/GUV-I 850-0 on safe working in laboratories (formerly

BGR/GUV-R 120 or ZH 1/119 laboratory guidelines issued by the employers’ liability insurance

association) (for Germany.
The message “TEMP. LIMIT” on the controller Display 2 indicates safety controller activity. If the buzzer

has been activated in the User Level (chap. 10), there is an audible alert. The safety controller controls the

chamber to the entered safety controller set-point until the temperature inside the chamber returns below

this temperature and until you then reset the alarm message with the EXIT button.
The safety controller set-point type and set-point can be set in the user level (chap. 10) of the program

controller.