2 programming with setting “step – BINDER KB 23 User Manual

Page 34

Advertising
background image

KB (E3.1 + E5.1) 11/2014

page 34/100


Program table corresponding to the diagram (with default setting “Ramp”):

Section


SEC

Temperature

set-point

[ °C]

TEMP

Section length

[hh.mm]

TIME

Fan speed

[%]

FAN

Operation lines

*

O.LINE

S01

40

00:30

50

000

S02

40

00:01

100

000

S03

60

01:30

100

000

S04

60

00:01

100

000

S05

80

01:00

100

000

S06

80

00:01

100

000

S07

20

03:20

100

000

S08

20

00:01

100

000

* Only with option zero-voltage relay outputs via operation lines, see chap. 16.4.
The values of such a program table can now be entered to the RD3 program controller (chap. 8.2).

The end point of the desired cycle must be programmed with an additional section (in our examples S05

for set-point ramp and S08 for set-point step) with a section time of at least one minute. Otherwise, the

program will stop one section too early because the program line is incomplete.

8.1.2 Programming with setting “step”

With setting “Step” selected, you don’t need to program the transition section in the Program Editor.

With setting “step” the controller will equilibrate only to constant temperatures; programming

ramps is no longer possible.

The set-points are maintained constant for the duration of a program section. At the start of each program

section, the unit heats up with the maximum speed in order to attain the entered set-point.

Program entry as set-point step (example):

W/°C

t/min.

S01 S02 S03 S04

0

20

40

60

80

100

30

120

180

380