KB (E3.1 + E5.1) 11/2014

page 38/100

Selecting the next program sections to be entered

Display 1 shows

e.g. 02

(section S02 selected)

Display 2 shows

alternating

P01: --- SEC.

Section S02 has already been created.

enter new set-points for the individual parameters with X/W.

CONTINUE X/W

or:

Display 1 shows

e.g. 02

(section S02 selected)

Display 2 shows

alternating

P01: --- SEC.

Section S02 has not yet been created.

enter set-points for the individual parameters with X/W

NEW SEC. X/W

Select the next section to be entered using

arrow keys

Display 1 shows

e.g. 12.5

C

(actual temperature set-point)

Display 2 shows

alternating

S02:TEMP 12.5

(variable: temperature in °C)

CONTINUE X/W

(information: go on with X/W)

Enter the temperature set-point of S02 in °C

using arrow keys

Etc.


If all sections up to S10 or up to S20 have been programmed, section S01 follows again. In order to quit

the entry mode, press the

EXIT button several times or wait 120 sec → the controller will then return to

Normal Display.

When changing the set-point, check the setting of the safety controller in the user level (chap.

10) if the safety controller has been set to “limit” mode.

If the fan is operated with less than 100 % speed, the temperature performance and the spatial

exactitude of the temperature can differ from the manufacturer’s specifications. The fan speed

rate should only be reduced if absolutely necessary to meet special requirements.

KB 23: Cooling power decreases when the fan speed is reduced to 20% or below.