F. SET RUN DURATION (Refer to Figure 8)

RUN DURATION

The Duration timer allows measurements runs to be performed over a specified time.
At the end of this time the measurement will automatically be stopped. From the
Menu select Settings (1). Use the cursor keys (C) to navigate to the Run Duration
icon (2). Press Select (B).

DURATION TIMER

Press (B) to edit the Duration timer to either On or Off and use cursors (C) to change
the selection. Press Save (B) to continue.

Figure 8 Run Duration (Sheet 1)

DURATION. Press cursor key (C) to navigate to the Duration field. Press Edit (B) to
change the Duration time. Use the Up/Down cursors (C) to make changes as
required. Range of duration/integration time is 00:00:01 to 24:00:00 in 1 second
steps.

Press Save (B) to save the changes. Press Exit (A) Settings screen.