9. APPENDIX B ADDITIONAL INFORMATION


IEC61672-1:2002 section 9.3 (Information for testing)

a) Reference Sound Pressure Level

The reference sound pressure level is selectable to be 94 dB or 114 dB.

b) Reference Level Range

The CEL-62X is a single level range instrument of 0 to 140 dB.

c) Microphone Reference Point

The microphone reference point is the centre of the diaphragm of the
microphone. The 0

o

reference direction is perpendicular to the microphone

diaphragm.

d) Acoustical Frequency Response Testing

The pressure to free field correction data for the CEL-251 microphone used for
periodic testing.

Frequency 0

o

Free

Field

Corrections

using a
Bruel &

Kjear 4226

Calibrator

0

o

Free

Field

Corrections

with Wind

Screen

using a
Bruel &

Kjear 4226

Calibrator

0

o

Free Field

Corrections

using a
Bruel &

Kjear

UA0033

Electrostatic

Actuator

0

o

Free Field

Corrections

with Wind

Screen using

a Bruel &

Kjear

UA0033

Electrostatic

Actuator

Expanded

Uncertainty

of

Corrections

at 95%

probability

(k=2)

Hz dB dB dB dB dB

31.5 -0.1 -0.1 -0.1 -0.1 0.2

63 -0.2 -0.2 -0.2 -0.2 0.2

125 -0.2 -0.2 -0.2 -0.2 0.2

250 -0.2 -0.2 -0.2 -0.2 0.2

500 -0.2 -0.2 -0.2 -0.2 0.2

1000 0 0.1

0

0.1 0.2

2000 0.3 0.7

0.3

0.7

0.3

4000 0.7 1.4

0.8

1.5

0.3

8000 2.8 2.5

3.1

2.8

0.4

12500 5.3

4.1

6.2

5.0

0.6

16000 6.4

4.5

7.8

5.9

0.6

Casella CEL Limited