6. SPECIFICATIONS

GENERAL

The CEL-62X provides SPL, Integrating and Octave band noise measurement
compliant with the following international standards:-

IEC 61672-1 2002-5 (Electro-Acoustics – Sound Level Meters) Group ‘X’
instruments. Performance of Class 1 or 2 as relevant to instrument model.

IEC 60651: 1979, IEC 60804: 2000, ANSI S1.4: 1983, ANSI S1

1/1 Octave and 1/3 Octave Filters comply with EN61260: 1996, Class 0 and ANSI
S1.11 1986, Order-3 Type 0C.

MEASUREMENT RANGE

Single measurement range up to 140.2dB(A) RMS and 143.3dB (C) Peak. Linearity
range from 10dB above noise floor.

RMS FREQUENCY WEIGHTINGS

A, C and Z filter weightings, satisfying IEC 61672-1: 2002 Class 1.

OCTAVE MEASUREMENT

Real-time 11 bands with centre frequencies from 16Hz to 16KHz. Displayed
Spectrum may be pre-weighted with A, C or Z.

Z weighted octaves only are stored to be post weighted in dB36 software.

PEAK MEASUREMENT

A, C and Z weighted from 65.0 to 143.3dB

RMS DETECTOR

Digitally derived True root-mean-square detection (RMS), 0.1dB display resolution.

NOISE FLOOR

Total inherent Noise including microphone thermal noise @20°C - <22dB(A) Class
1, <30dB(A) Class 2.

Electrical Noise floor <20dB(A).

FREQUENCY RESPONSE

6Hz to 20kHz (upper and lower 3dB frequencies). Digital Sampling rate 67.2KHz

Overall Class 1 and 2 frequency response compliant with IEC 61672-1: (2002)

TIME WEIGHTINGS

Fast, Slow and Impulsive according to IEC 61672-1: (2002).

CORRECTION FILTERS

Built in correction filter for random incidence sound pressure fields.

REFERENCE DIRECTION

For free field measurements the reference direction is perpendicular to the
microphone diaphragm.

REFERENCE CONDITIONS

23

o

C air temperature, 50% Relative Humidity,

101.325kPa atmospheric pressure.

Nominal reference level = 114.0dB at 1kHz