MEASUREMENT VIEW.

Predefined views are available to satisfy measurements for international workplace
noise legislation. Use the cursor keys (C) to select the required view and press (B) to
activate the measurement view. The Measurement View with the word ACTIVE next
to it will be used for measurement runs. To view the details press View (B).

On CEL-621 units, two fixed environmental views are available which provide
additional statistical parameters (L

n%

) and parameters unique to workplace noise are

removed.

USER 1 AND 2 SETTINGS.

Two customiseable views are available in which measurement weightings and
paramaters can be set. When one of these views is Active by pressing the (B) key,
press (B) again to edit the view. Left right cursor keys are used to make changes
and up/down cursor keys are used to move to the next field. On this screen you can
define the following:

RESPONSE (Random/Freefield)

Random microphone response is typically used to conform to US legislation for
workplace noise whereas Freefield is used throughout the EU.

LAVG THRESHOLD (70-90dB)

The threshold is used for the calculation of L

AVG

, which is the level at which all noise

below is ignored. Note with threshold set to zero the threshold is off.

OCTAVE WEIGHTING (A, C or Z)

The frequency weighting used to display 1/1 or 1/3 octave band results. Note that
regardless of the settings selected, Z weighted measurements are stored so the data
can be post-processed on a PC to C or A weighting.

OCTAVE TIME CONST (F or S).

The time weighting used to display 1/1 or 1/3 octave band results. Regardless of the
selection, both time weightings are stored simultaneously for L

MAX

results. Time

weightings of Fast (F) or Slow (S) do not apply to L

eq

measurements.

Press (D) to restore defaults. Press (B) to open the Functions screens for the
selected ‘User 1 View’.

Figure 4 Setting Options (Sheet 1)