Remove the acoustic calibrator. Press EXIT (A) to go back to the Stop screen. Press
and hold the acoustic calibrator ON/OFF Key to switch the calibrator off.

Figure 15 Instrument Calibration

L. EXTRACTING DATA (Refer to Figure 16)

Each measurement run is stored as a file within the CEL-62X in a .CSV format. This
format will automatically open with MS EXCEL or similar office applications. The files
are named consecutively from 1 to 100, so the first run will be named R001.CSV. All
measurement parameters are contained within these files, regardless of the
displayed parameters selected within the instrument setup.

The CEL-62X acts as a removable storage device when connected to a PC by the
CMC51 cable. See Figure 18 for the location of the USB connection on the CEL-600
Series. Once connected, the PC will automatically detect the instrument is
connected and within a few seconds have installed the necessary drivers. An
Explorer window will open on the PC to show the files on the instrument. Simply
copy the files to your PC as required.