OPERATING ENVIRONMENTAL CONDITIONS

Humidity: 5 to 90%RH in the absence of condensation.

Temperature range: -10 to +50

o

C (Class 1), 0 to +40

o

C (Class 2)

Pressure: 65 to 108kPa

EFFECT OF TEMPERATURE

Electrical stability of instrument < ±0.2 dB over the range -10 to +50

o

C

EFFECTS OF HUMIDITY

Less than ±0.5dB over the range 25 to 90% relative humidity (non-condensing),
relative to the value at reference conditions.

STORAGE ENVIRONMENTAL CONDITIONS

Humidity: 0 to 90%RH in the absence of condensation.

Temperature: -20 to +60

o

C.

Pressure: 65 to 108kPa

MICROPHONE

Class 2: CEL-252 ½” Class 2, 30mV/Pa pre-polarised back electret.

Class 1: CEL-251 ½” Class 1, 50mV/Pa pre-polarised back electret.

Note: Instrument has fixed preamp so class is determined by microphone only.

CALIBRATION

Auto calibration by application of 1KHz calibrator, nominal level 114 or 94dB +/-1dB.

Auto calibrated to a user specified reference level with recording of date, time and
offset.

POWER SUPPLY

External DC: 9 to 14V DC at typically 170mA. 2.1mm Power Connector.

Batteries: 3x AA Alkaline or rechargeable cells

Battery Life: Typically >8 hours in Broadband mode with backlight continuously on
low intensity. High backlight setting will have a detrimental affect on battery life.

12 hours measuring with backlight off.

INTERNAL CLOCK:

Date and time accuracy better than 2 sec per day.

LANGUAGES:

English (default), Spanish, French, German, Italian, Portuguese, Chinese, and
Brazilian Portuguese.

ELECTROMAGNETIC COMPATIBILITY

Instrumentation is designed and tested to comply with the following EMC and ESD
Standards:

IEC 61000-4-2 Testing and Measuring Techniques - Electrostatic discharge immunity
tests.

IEC 61000-4-3 Electromagnetic compatibility (EMC) - Radiated electromagnetic field
tests.

IEC 61000-4-6 Electromagnetic compatibility (EMC) - Immunity to conducted
disturbances induced by radio frequency fields. Tested at 10V/m or greater.

EFFECTS OF AC POWER FREQUENCY FIELDS

Less than ±0.5dB change from 74dBA 925Hz reference level when subjected to
160A/m AC magnetic field at 50 and 60Hz.