Figure b-4. properties, settings tab – Basler Electric BE1-79A User Manual
Page 69
9310200990 Rev J
BE1-79A Terminal Communication
B-3
Step 7: Click File/Properties on the menu bar. Click the Settings tab.
a. Make the following selections using Figure B-5 as a guide:
Check the Terminal Keys radio button.
Select VT-100 emulation.
Set Backscroll buffer lines to the maximum setting of 500.
Figure B-4. Properties, Settings Tab
b. Click the ASCII Setup button. Make the following selections using Figure B-5 as a guide:
Place a check at Send line ends…
ASCII Sending
Place a check at Echo typed characters…
Set the Line delay setting to 100 milliseconds.
Set the Character delay setting to 0 milliseconds.
Disable Append line feeds…by leaving the box unchecked.
ASCII Receiving
Disable Force incoming… by leaving the box unchecked.
Place a check at Wrap lines…
c. Click “OK”.
d. Click “OK”.
Step 8: Click File and click Save.
NOTE
Settings changes do not become active until the settings are saved.