Cleaning the Sensor Assembly

Note, in some cases, contamination of the sensor is the only cause of the
additional heat transfer during the “no flow” test. Remove the sensor, and
clean the sensor (use an appropriate non-corrosive solvent to remove the build-up). A soft brush can
be used to gently clean the sensing surface, using caution to avoid damaging the sensor elements.
Once clean, re-install the sensor as per the instruction manual.

Contact ONICON Service for assistance if the mW readings differ by more than ± 10mW.

CLIMATE ACTION RESERVE LANDFILL PROJECT PROTOCOL VER. 4.0 JUNE 29, 2011

ONICON’s zero flow calibration test complies with the requirements of section 6.2 for field checking
the flow sensor for calibration accuracy. Percent drift can be determined by multiplying the mW
change from factory value (see 2) by 1.0% (i.e. each mW change equals 1% drift).