Flow rate totalizer, Setup, Maintenance – Super Systems eFLO User Manual

Page 14: Flow rate totalizer setup, He setup, N below

Advertising
background image

eFlo Electronic Flow Meter Operations Manual

Super Systems Inc.

Page 14 of 25

Flow Rate Totalizer

The eFlo meter’s flow rate totalizer records cumulative flow rates measured by the meter. The

totalizer is active at all times. The maximum totalized value is 999,999,999,999 (the unit of

measurement is any unit of flow measurement). To show the totalized value, press and hold the

Totalizer button. The totalized value will begin scrolling across the LED screen if the value is

greater than what the screen can display.

To reset the totalizer to zero, press and hold the Reset button for five seconds.

Setup

The Setup button will open the Setup menu when the button is held for five seconds. The options

are displayed on the LED screen. Scroll through the options by pressing the Auto/Manual

button. Change the values by pressing the Up and Down buttons.

The available options are as follows.

GAIN: Gain is used to control the speed of valve response for automatic control. The higher the

value, the more quickly the valve will open and close. The lower the value, the more slowly it

opens and closes. The range for this value is 10 to 500.

DEAD (Deadband): Deadband is used to determine the area above and below the setpoint where

no control action will occur. If the deadband is too small, the control may oscillate around

setpoint. If the deadband is too high, the flow will not reach setpoint. The range for this value is

1 to the full scale flow value.

ADDR (Address): The Address is the RS485 Modbus address of the eFlo unit. The range for this

value is 1 to 249.

ZERO: The Zero value is the adjusted mA input from the pressure transducer. This value allows

the user to adjust the pressure transducer, which transmits a 4-20mA signal to indicate flow.

When no flow is present, the ideal value for the Zero value is 4.00mA (although the value will not

always be exactly 4.00). To change the Zero value, use the Up and Down buttons. To reset the

Zero value and remove any previous adjustments, press the Up and Down buttons

simultaneously. For more information, refer to the Zero Calibration section on page 15.

BAUD: The Baud is the serial baud rate. This value can be 9600 or 19200 (1920 represents

19200).

Maintenance

SSi recommends the following maintenance guidelines for the eFlo meter.

Clean any carbon or water from pressure transducer gas lines as needed. To do this:

1.

Remove the rubber tubing from one side of the pressure transducer located in the

electronics enclosure. (Note that the two tubes plumbed to the pressure transducer

must not be switched; otherwise, the meter will not read properly.) The other end of the

tubing should be connected to the stainless steel tubing plumbed vertically along the

right side of the meter.

2.

Apply air to the tubing to “flush” out any carbon and/or moisture, which may have

developed. Do not apply pressure to the pressure transducer.

3.

Reassemble the tubing to the pressure transducer port and repeat the procedure for the

other side of the pressure transducer.