5.

Using the manometer, measure the actual differential pressure at the selected flow

rate. Compare the differential pressure to the differential pressure shown on the

Calibration Report.

NOTE: Typically, a 0.5”wc deviation equates to 5% flow error.

6.

If the actual differential pressure is within acceptable limits (±4% of the full scale value),

no action is necessary. If the actual differential pressure is not within acceptable limits,

adjust the zero/span potentiometers inside the differential pressure transducer. The

pressure transducer is located inside the black enclosure on top of the unit.