2 guidelines for conduit, 3 guidelines for cable gland, 4 wiring connections to sensor – Emerson IFT9701 User Manual

Page 24: Guidelines for conduit, Guidelines for cable gland, Wiring connections to sensor, See section 3.4.2 for, Allations. see section 3.4.3 for information, Inals, see section 3.4.4, Warning

Advertising
background image

16

Model IFT9701 Transmitter Instruction Manual

Remotely Mounting the Transmitter

continued

3.4.2

Guidelines for conduit

If sealed metallic conduit is installed, it must provide 360° termination shielding for the enclosed
flowmeter cable.

1. Install a drip leg in conduit to prevent liquids from entering the junction box.

2. Connect the sealed end of the conduit to the ¾-inch NPT female conduit opening on the sensor

junction box.

3.4.3

Guidelines for cable gland

1. Prepare cable ends and assemble the supplied cable gland according to the instructions that are

enclosed with the cable preparation kit.

2. Connect the ¾-inch NPT male cable gland to the ¾-inch NPT female conduit opening on the

sensor junction box.

3.4.4

Wiring connections to sensor

1. At the sensor, insert the ends of the individual wires into the terminals inside the sensor

junction box. Match the wire colors of the cable with the wire colors at the sensor wiring
terminal as described in Figure 3-7 and Table 3-3. No bare wire should remain exposed.

2. If possible, position the junction box on the sensor so the conduit entrance points downward to

prevent moisture from entering.

3. At the transmitter, if cable is not preinstalled, connect the cable to the transmitter terminals as

indicated in Figure 3-8.

WARNING

Failure to seal sensor junction box or transmitter
housing could cause a short circuit, which would
result in measurement error or flowmeter failure.

• Seal all conduit openings.
• Install drip legs in cable or conduit.
• Fully tighten sensor junction box and transmitter housing

covers. See Figure 3-8.

Advertising