Model IFT9701 Transmitter Instruction Manual

27

Power Supply and Output Wiring

continued

Ge
tti
n

g

Sta

rted

P

o

wer

Su
pp
ly

an

d O

u

tp
ut Wi

ri

ng

Mounting the

Remo

te T

ransmitter

Be
fo

re
Y

o

u

Begin

Figure 4-8

Wiring to Series 3000 applications platform with I/O cable

Figure 4-9

Wiring to Series 3000 applications platform with screw-type or solder-tail terminals

Figure 4-10 Wiring to field-mount Series 3000 applications platform

(1) Paired cables with shield. Maximum length is 50 feet for 28 AWG wire (15 meters for 0,1 mm² wire),

500 feet for 22 AWG wire (150 meters for 0,3 mm

2

wire).

(2) Do not connect shields at this end.

IFT9701 pulse

output terminals

Series 3000 I/O cable

terminals

See Note 2

Earth

ground

(1) Paired cables with shield. Maximum length is 50 feet for

28 AWG wire (15 meters for 0,1 mm² wire), 500 feet for 22 AWG
wire (150 meters for 0,3 mm

2

wire).

(2) Do not connect shields at this end.

IFT9701 pulse

output terminals

Series 3000 screw-type or

solder-tail terminals

See Note 2

Earth

ground

(1) Paired cables with shield. Maximum length is 50 feet for 28 AWG wire (15 meters for 0,1 mm² wire),

500 feet for 22 AWG wire (150 meters for 0,3 mm

2

wire).

(2) Do not connect shields at this end.

IFT9701 pulse

output terminals

Series 3000 field-mount

terminals

See Note 2

Earth

ground