Model IFT9701 Transmitter Instruction Manual

43

Configuration with a HART Communicator

continued

Co
nfig

uratio

n wit

h

Co
mm

u

n

ic

ato

r

T

ro

u

b

lesho

ot
ing

Con

fig

ur
atio

n

w
ith
Pr
oL
in

k

II

Flo

w

meter Star

tup

To calculate the flow calibration factor:

1. Select a mass flow unit, and send the unit to the flowmeter memory. See Table 6-1.

2. To enter the flow calibration factor into flowmeter memory, refer to the menu below and

follow these steps:

a.

At the online menu, choose 4 (detailed setup).

b.

At the detailed setup menu, choose 1 (characterize sensor).

c.

At the characterize sensor menu, choose 1 (flow calibration factor).

d. Enter the 8-digit flow calibration factor listed on the flowmeter serial number tag. The

flow calibration factor can also be found on the certificate that is shipped with the meter.

e.

When the display returns to the characterize sensor menu, press F2 (send) to send the
factor to the flowmeter memory.

3. Zero the flowmeter (see Section 6.3.1).

4. Refer to the menu below, and follow these steps:

a.

Reset the internal totalizer to zero.

b.

Run three batches of fluid, resetting the scale and totalizer between batches. For each
batch, record the weights indicated by the scale and the totalizer.

5. Divide Total Weight

scale

by Total Weight

totalizer

. This is the mass-flow meter factor. Record the

meter factor.

1 Process variables
2 Diag/service
3 Basic setup
4 DETAILED SETUP
5 Review

1 CHARIZE SENSOR
2 Config fld dev var
3 Config outputs
4 Device information

1 FLOW CAL
2 Density cal factr

1 View fld dev vars
2 View output vars
3 View status
4 TOTALIZER CNTRL

1 Totl
2 Reset totalizer

1 PROCESS VARIABLES
2 Diag/Service
3 Basic setup
4 Detailed setup
5 Review

Weight

scale

Weight

totalizer

First batch

Second batch

Third batch

Total

Mass-flow meter factor