2 optional lcd, 3 fault outputs, Optional lcd – Emerson IFT9701 User Manual

Page 66: Fault outputs

Advertising
background image

58

Model IFT9701 Transmitter Instruction Manual

Troubleshooting

continued

Figure 8-1

Transmitter terminals and diagnostic LED

8.3.2

Optional LCD

Except for ATEX Zone 1 areas, the transmitter is available with an optional LCD. The 5-digit LCD is
on the cover of the field wiring compartment, and indicates the mass or volume flow rate or the
operating condition of the flowmeter. Table 8-2 explains the operating conditions indicated by the
LCD.

8.3.3

Fault outputs

The transmitter produces downscale or upscale outputs to indicate a fault.

Downscale: The milliamp output goes to 2 mA; the pulse output goes to 0 Hz.

Upscale: The milliamp output goes to 22 mA; the pulse output goes to 7200 Hz.

Unless otherwise specified on the order, jumpers are set so the transmitter generates downscale fault
outputs. To change jumper settings for fault outputs, see Section 2.3.

Table 8-1

Normal conditions indicated by LED

Diagnostic LED does this:

Condition

Blinks ON once per second (25% ON, 75% OFF)

Normal operation

Remains ON

Startup and initialization, zero in progress, transmitter hardware
failure

Blinks OFF once per second (75% ON, 25% OFF)

Slug flow (density below or above user-defined limits)

Blinks ON 4 times per second

Fault condition

Remains OFF

Power-supply failure, transmitter hardware failure

Power supply
• If transmitter fails to

produce outputs,
check for reversed
polarity

• Make sure actual

power supply
matches power
supply shown on
label

• AC power applied to

DC transmitter will
damage transmitter

• Power-supply fuses

are not replaceable

Ground screw

for power supply

Case ground

Diagnostic LED

Shields (GND) for wiring to sensor
(remote-mount transmitters only)

Sensor wiring
compartment
(intrinsically safe)

Pulse output
terminals

Milliamp output

terminals

+

20–30V

or

N

L

85–250V~

Intrinsically safe
terminal block

Advertising