To check integrity of sensor wiring circuits:

1. Disconnect the transmitter's power supply.

2. Disconnect sensor wiring from the transmitter's intrinsically safe terminal block in the sensor

wiring compartment. See Figure 8-1.

3. Use a DMM to measure resistance between wire pairs, as indicated in Table 8-4.

4. If the transmitter is remotely mounted from the sensor, repeat the measurements at the

terminal block in the sensor junction box to distinguish cable failure from sensor failure.

8.6

Over range and sensor failure conditions

If a sensor failure occurs, if the flowmeter cable is faulty, or if flow, temperature, or density goes
outside the sensor limits, all the following occur:

The transmitter produces fault outputs.

The diagnostic LED blinks ON 4 times per second.

The optional LCD reads “SENS0”.

Table 8-3

Sensor terminal designations

Terminal number

Wire color

Function

1

Brown

Drive +

2

Red

Drive –

3

Orange

Temperature –

4

Yellow

Temperature lead length compensator

5

Green

Left pickoff +

6

Blue

Right pickoff +

7

Violet

Temperature +

8

Gray

Right pickoff –

9

White

Left pickoff –

Table 8-4

Normal resistance and voltage ranges for flowmeter circuits

Notes

• Temperature sensor value increases 0.38675 ohms per °C increase in temperature.
• Nominal resistance values will vary 40% per 100 °C. However, confirming an open coil or shorted coil is more important

than any slight deviation from the resistance values presented below.

• Resistance across terminals 6 and 8 (right pickoff) should be within 10% of resistance across terminals 5 and 9 (left

pickoff).

• Resistance values depend on the sensor model and date of manufacture.

Circuit

Wire colors

Sensor
terminals

Nominal resistance range

Drive coil

Red to brown

1 to 2

8 to 2650

Left pickoff

Green to white

5 to 9

15.9 to 300

Right pickoff

Blue to gray

6 to 8

15.9 to 300

Temperature sensor

Orange to violet

3 to 7

100

at 0°C + 0.38675

/°C

Lead length compensator

Yellow to violet

4 to 7

100

at 0°C + 0.38675

/°C

Advertising