3 calibration procedures, 1 auto zero, Calibration procedures 7.3.1 – Emerson IFT9701 User Manual

Page 59: Auto zero

Advertising
background image

Model IFT9701 Transmitter Instruction Manual

51

Configuration with ProLink II Software

continued

Co
nfig

uratio

n wit

h

Co
mm

u

n

ic

ato

r

T

ro

u

b

lesho

ot
ing

Con

fig

ur
atio

n

w
ith
Pr
oL
in

k

II

Flo

w

meter Star

tup

To configure the pulse output:

1. From the ProLink menu, click on Configuration.

2. Click on the Frequency tab. The Frequency panel appears, as shown in Figure 7-3.

Figure 7-4

Configuration window – Frequency panel

3. The pulse output produces a frequency proportional to the flow rate. Pulse output scaling

requires entry of frequency and flow rate values. To scale the pulse output:

a.

In the text box labeled Freq Factor, enter a value for the frequency in pulses per second
(Hertz) that corresponds to the maximum flow rate.

b.

In the text box labeled Rate Factor, enter the maximum value of the flow rate. The entered
value corresponds to the frequency that was entered in the Freq Factor text box.

7.3

Calibration procedures

Calibration procedures include zeroing and flow calibration.

7.3.1

Auto zero

Flowmeter zeroing establishes flowmeter response to zero flow and sets a baseline for flow
measurement.