Model IFT9701 Transmitter Instruction Manual

Configuration with a HART Communicator

continued

6.2

Configuration parameters

Use the basic setup menu to perform the following tasks:

Assigning a HART tag to the flowmeter

Changing measurement units for the mass or volume flow rate

Setting range values for the milliamp output

Scaling the pulse output

Use the detailed setup menu to perform the following tasks:

Changing measurement units for density or temperature

Changing the flow cutoff value

Changing the internal damping value

Changing the flow direction parameter

6.2.1

HART tag

The HART tag consists of up to eight characters that identify the flowmeter when it communicates
with other devices in a HART multidrop network.

To assign a HART tag to a flowmeter, use the menu below and follow these steps:

1. At the online menu, choose 3 (basic setup).

2. At the basic setup menu, choose 1 (tag).

3. Enter the desired tag of up to eight characters, then press F4 (enter). The tag can include

spaces and periods.

6.2.2

Measurement units

The meter can use one mass flow unit or one volume flow unit. The meter can also provide density
and temperature indication to a HART Communicator.

To select a measurement unit for mass flow or volume flow, use the menu below and follow these
steps:

1. At the online menu, choose 3 (basic setup).

2. At the basic setup menu, choose 2 (primary variable unit).

3. Use the down arrow (Ø) or up arrow (×) to select the desired measurement unit, then press F4

(enter). Available measurement units for mass flow and volume flow are listed in Table 6-1.

1 Process variables
2 Diag/service
3 BASIC SETUP
4 Detailed setup
5 Review

1 TAG
2 PV unit
3 Analog range vals
4 Freq factr

1 Process variables
2 Diag/Service
3 BASIC SETUP
4 Detailed setup
5 Review

1 Tag
2 PV UNIT
3 Analog range vals
4 Freq factr