A.2.2 local display (optional), A.2.3 low-flow cutoff, A.2.4 slug-flow inhibit – Emerson IFT9701 User Manual

Page 74: A.2.5 damping, A.2.6 fault indication, A.2.7 output testing, A.2.8 power supply options, A.2.2, Local display (optional), A.2.3

Advertising
background image

66

Model IFT9701 Transmitter Instruction Manual

IFT9701 Specifications

continued

A.2.2

Local display (optional)

5-digit, alphanumeric, liquid crystal display (LCD) installed on the field-wiring compartment cover.
The LCD shows flow rate, and indicates slug flow, flowmeter zeroing in progress, and electronic
faults.

Note: Not available for ATEX Zone 1 applications.

A.2.3

Low-flow cutoff

Flow rate below cutoff causes outputs to default to level that indicates zero flow and totalizer stops
counting

A.2.4

Slug-flow inhibit

When transmitter senses density outside user-selected limits, outputs default to levels indicating zero
flow and totalizer stops counting

A.2.5

Damping

User-selected time constant of 0.1, 0.2, 0.4, 0.8, 1.6, 3.2, 6.4, or 12.8 seconds

A.2.6

Fault indication

Fault condition can be indicated by user-selected downscale (2 mA, 0 Hz) or upscale (22 mA,
7200 Hz) output levels, Bell 202 communications, and optional display

A.2.7

Output testing

Current source

Can output a specified current between 2 and 22 mA

Frequency source

Can output a specified frequency between 0.01 and 1200 Hz

A.2.8

Power supply options

85 to 250 VAC

45 to 65 Hz, 9 watts maximum power dissipation

20 VA maximum apparent line power

Fused at 250 V/0.4 A IEC delayed, nonreplaceable

Meets low-voltage directive 73/23/EEC