Warning – Amprobe Multitest-1000 Continuity-Tester User Manual

Page 14

Advertising
background image

CABLES USED FOR THE TEST

If the cables are changed or extended be sure and
recalibrate the instrument. During a continuity
test, if the resistance value free of calibration (that
is the resistance value less the calibration offset

value) is negative, the symbols

ν

as well as

blinking CAL are displayed (refer to 5

th

screen

paragraph 4.1.3).

WARNING

Never disconnect the test leads while the word “Measuring” is being
displayed.


4.1.1.1. Procedure to cancel calibration parameters and cal symbol

LOW

ν>

5

cal

Measuring

To cancel the calibration parameters (and
the symbol CAL) it is necessary to perform a
calibration procedure with a resistance
greater than 5

Ω (for example with the test

leads disconnected). When a cancellation is
completed this screen is displayed first,
followed by the screen below:

Message >5

Ω:

means that the
instrument
detected a
resistance greater
than 5

Ω therefore

it will proceed
with Reset
procedure.


LOW

es

cal

The screen alongside is displayed for 2
seconds
, after which the instrument emits a
long beep
and then displays the default
screen relative to the LOW

Ω test under

AUTO mode without the symbol CAL.

Message rES:
means that the
instrument
cancelled
(RESET) the
calibration
parameter