4.2.2. Special cases which may occur during the tests "MAN" & "TIMER"

)

R

ISO

> 999

M

500 V 20

s


MAN

In case a value of R

ISO

greater than

R

MAX

was detected (depending on the

selected voltage see following Note
table, the instrument emits a double
beep
at the end of the test indicating
the positive outcome of the test
and
displays a screen similar to the screen
alongside.

The symbol ">"
means that the
resistance value
R

ISO

is greater

than R

MAX

Maximum resistance value that can be measured
(999

Ω is displayed if a rated voltage of 500V was

selected see table).

Selected mode
MAN

Test duration

)

R

ISO

ν0.01

M

500 V 20

s


MAN

In case a test is performed at a voltage
lower than the set rated voltage
, at
the end of the test the instrument emits
a long beep and displays a screen
similar to the screen alongside.

Selected mode
MAN

ATTENTION: the test of resistance R

ISO

was

performed at a voltage value lower than the set
rated voltage
. Low insulation case. This case
occurs under low insulation conditions or in the
presence of capacitance on the installation.

SAVE

SAVING:

The test can be stored pressing the SAVE key twice (refer to
paragraph 5.1).


)

R

ISO

ν 33V

VOL

AG

If the terminal voltage is greater than
30V, the instrument does not perform
the test, displays the screen alongside
for 5 seconds after which, it shows the
default screen relative to the R

ISO

test

under selected mode.

ATTENTION: the
test was not
carried out.
Check that the
circuit is not
energized.

SAVE

THIS RESULT CANNOT BE SAVED.