Warning – Amprobe Multitest-1000 Continuity-Tester User Manual

Page 20

Advertising
background image

4.2.1. Procedure for measuring insulation resistance in any mode
1. Select the desired mode by means of the FUNC key.
2. Connect the black and blue cables to the instrument input terminals T1 and T4

respectively.

B2

B3

B4

B1

M

Insulation between phase and earth in an electrical

installation using untied cables.


3. If the cables supplied with the instrument are not long enough for the measurement you can

extend the blue cable.

4. Connect the instrument terminals to the object that is to be submitted to the insulation

test after disconnecting the circuit under test and all the relative loads (see
previous picture).


5.

By pressing the U

n

key select the test voltage suitable for the type of

test to be carried out (see table). The values to be selected are:
• 50V (test on telecommunication system)
• 100V

• 500V

• 250V

• 1000V

Standard

Brief description

Test voltage

Maximum limit value

CEI 64-8/6

Systems SELV or PELV

Systems up to 500V (Civil installations)

Systems over 500V

250VDC
500VDC

1000VDC

> 0.250M

> 0.500M

> 1.0M

CEI 64-8/4

Floor and wall insulation in civil installations

Floor and wall insulation in systems over 500V

500VDC

1000VDC

> 50k

Ω (if V<500V)

> 100k

Ω (if V>500V)

EN60439

Electrical panel boards 230/400V

500VDC

> 230k

EN60204

Electrical equipment of machines

500VDC

> 1M

CEI 64-4

Floor insulation in medical rooms

500VDC

<1M

Ω (if the floor is at least 1 year old)

<100M

Ω (if the floor is at least 1year old)

Test voltage values and the relative maximum limit values for the most common kinds of test.

WARNING

If “Measuring” is displayed the instrument is performing the measurement.
During this phase do not disconnect the test leads, as the circuit under test
may remain charged at a dangerous voltage due to the parasite capacitance