ANOMALOUS CASES WHICH MAY OCCUR DURING EARTH

ρ TESTS

)

- - -

- - -

c


If the amperometric circuit is
interrupted, when pressing
START/STOP the instrument can’t read
the minimum current, therefore a
screen similar to the one beside
appears. Make sure that the terminals
are connected correctly and that the
amperometric rod (blue conductor) has
not been driven into non-conductive
ground. If necessary, pour water
around the rod.

rC indicates a
high resistance
value.



SAVE

THE PREVIOUS RESULTS CANNOT BE SAVED.


)

31

V

VOL

AG

If the instrument measures an
interfering voltage greater than 30V
on the amperometric circuit
, it
doesn’t perform the test and a screen
similar to the one beside appears.

Voltage value on the amperometric circuit


SAVE

THE PREVIOUS RESULT CANNOT BE SAVED.