Mode "auto, Mode "r+timer" or “r-timer – Amprobe Multitest-1000 Continuity-Tester User Manual

Page 16

Advertising
background image

4.1.2.1. Mode

"AUTO"

)

CAL

LOW

.07

219

mA

AUTO

At the end of the test, if the average
resistance value Ravg
is less than 5

the instrument emits a double beep
indicating the positive outcome of the
test
and displays a screen similar to the
screen alongside.

Average test
current value
Iavg.

Average
resistance value
Ravg.

SAVE

SAVING:

The test results can be stored pressing the SAVE key twice
(refer to paragraph 5.1).



4.1.2.2. Mode "R+TIMER" or “R-TIMER"

)

CAL LOW

.08

219

mA

5

s

R+
R-

TIMER

If during the test a resistance value
R+Timer
or R–Timer less than 5

was

detected, the instrument (after the set
time has elapsed) emits a double beep
indicating a positive outcome
of the
test & displays a screen similar to the
screen alongside.

Duration time of
the test.

Value of the test
current I+ or I-.

The symbols R

+

or R

-

are displayed.

SAVE

SAVING:

The test can be stored pressing the SAVE key twice (refer to
paragraph 5.1).