Neck – Siemens SOMATOM SENSATION CARDIAC A60 User Manual

Page 101

Advertising
background image

101

Neck

AngioCarotid

2

nd

Reconstr.

kV

120

Effective mAs

120

Slice collimation

0.75 mm

Slice width

5.0 mm

1.0 mm

Feed/Rotation

12.0 mm

Rotation time

0.5 sec.

Kernel

B30f

B30f

Increment

5.0 mm

0.7 mm

Image order

cr-ca

CTDI

w

9.4 mGy

Effective dose

Male: 1.6 mSv
Female: 1.7 mSv

CARE Bolus may be used to optimize the bolus timing.

Set the ROI for monitoring scan in the aortic arch
with triggering threshold of 120 HU, or use manual
triggering.

MIPthin: AngioCarotid

coronal

Image thickness

3

Distance between images

3

Number of images

20

For the 2

nd

reconstruction the Autoload into MIPthin

Range on the 3D Card is activated. The images will be
automatically loaded into 3D, MIPthin, and a coronal
MIPthin Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.

Contrast medium IV injection
Start delay

15 sec.

Flow rate

4 ml/sec.

Total amount

100 ml