Heartview ct pulmonaryecg – Siemens SOMATOM SENSATION CARDIAC A60 User Manual

Page 49

Advertising
background image

49

HeartView CT

PulmonaryECG

Indications:

This is a spiral scanning protocol with an ECG gating
technique for aortic and pulmonary studies, e. g. aortic
dissection or pulmonary emboli.

PulmonaryECG

kV

120

Effective mAs

200

Slice collimation

1.5 mm

Slice width

3.0 mm

Feed/Rotation

5.6 mm

Rotation time

0.42 sec.

Temporal resolution

Up to 105 ms*

Kernel

B30f

Increment

2.0 mm

Image order

cr-ca

CTDI

w

14.4 mGy

Effective dose

Male: 5.7 mSv
Female: 7.3 mSv

Topogram:
AP, 512 mm.
From the aorta arch
until the tip of the
sternum.
A typical range of
30 cm can be covered
in 23 s.

* Depends on heart rate.