Head – Siemens SOMATOM SENSATION CARDIAC A60 User Manual

Page 93

Advertising
background image

93

Head

AngioThinSlice

2

nd

Reconstr.

kV

120

Effective mAs

130

Slice collimation

0.75 mm

Slice width

4.0 mm

0.75 mm

Feed/Rotation

6.8 mm

Rotation time

0.5 sec.

Kernel

H20f

H10f

Increment

4.0 mm

0.5 mm

Image order

ca-cr

CTDI

w

27.4 mGy

Effective dose

Male: 0.6 mSv
Female: 0.7 mSv

Use of CARE Bolus with monitoring scans positioned at
the level of the basilar artery or carotid artery. Set the
trigger threshold at 120 HU, or use manual triggering.

MPRthick: AngioThinSlice

coronal

Image thickness

3

Distance between images

3

Number of images

35

For the 2

nd

reconstruction the Autoload into MPRthick

Range on the 3D Card is activated. The images will be
automatically loaded into 3D, MPRthick, and a coronal
MPRthick Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.

Contrast medium IV injection
Start delay

18 sec.

Flow rate

3.5 ml/sec.

Total amount

75 ml

Advertising