Heartview ct coronarycare – Siemens SOMATOM SENSATION CARDIAC A60 User Manual

Page 40

Advertising
background image

40

HeartView CT

CoronaryCARE

Indications:

This is a spiral scanning protocol, using a rotation
time of 0.42 s, ECG gating technique and a dedicated
cardiac filter which can reduce images noise thus
makes the dose reduction possible for coronary CTA
studies.

Topogram:
AP, 512 mm.
Approximately, from
the carina until the
apex of the heart.
A typical range of
12 cm covering the
entire heart can
be done in 18.7 s.

CoronaryCARE

kV

120

Effective mAs

267

Slice collimation

0.75 mm

Slice width

1 mm

Feed/Rotation

2.8 mm

Rotation time

0.42 sec.

Temporal resolution

Up to 105 ms*

Kernel

B30f

Increment

0.5 mm

Image order

cr-ca

CTDI

w

22.4 mGy

Effective dose

Male: 3.8 mSv
Female: 5.5 mSv

* Depends on heart rate.

Advertising