Heartview ct, Cascorespistd0.5s – Siemens SOMATOM SENSATION CARDIAC A60 User Manual

Page 33

Advertising
background image

33

HeartView CT

* Depends on heart rate.

CaScoreSpiStd0.5s

Indications:

This is a spiral scanning protocol, using an ECG gating
technique for coronary calcium scoring studies, with
a rotation time of 0.5 seconds.

Topogram:
AP, 512 mm.
From the carina until
the apex of the heart.
A typical range of
15 cm covering the
entire heart can
be done in 11.7 s.

CaScoreSpiStd0.5s

kV

120

Effective mAs

133

Slice collimation

1.5 mm

Slice width

3 mm

Feed/Rotation

6.8 mm

Rotation time

0.5 sec.

Temporal Resolution

Up to125 ms*

Kernel

B35f

Increment

1.5 mm

Image order

cr-ca

CTDI

w

9.6 mGy

Effective Dose

Male: 2.1 mSv
Female: 3.1 mSv