LungCare

LungCare

kV

120

Effective mAs

20

Slice collimation

0.75 mm

Slice width

1.0 mm

Feed/Rotation

18.0 mm

Rotation time

0.5 sec.

Kernel

B50f

Increment

0.5 mm

Image order

cr-ca

CTDI

w

1.7 mGy

Effective dose

Male: 0.6 mSv
Female: 0.7 mSv

A dedicated low dose Spiral mode for the syngo Lung
CARE evaluation.

Indications:

Lung studies with low dose setting, e. g. early visuali-
zation of pulmonary nodules.

A typical thorax study in a range of 30 cm will be
covered in 8.9 s.

We recommend using a tube voltage of at least
120 kV.

For further information on the syngo Lung CARE
Application, please refer to the Application Guide
“Clinical Options”.