Children abdcta – Siemens SOMATOM SENSATION CARDIAC A60 User Manual

Page 86

Advertising
background image

86

Children

AbdCTA

Indications:

For abdominal CT Angio studies, e. g. vascular
abnormalities, aneurysms, etc.

AbdCTA

2

nd

Reconstr.

kV

80

Effective mAs

*

Slice collimation

0.75 mm

Slice width

3.0 mm

1.0 mm

Feed/Rotation

12 mm

Rotation time

0.5 sec.

Kernel

B30f

B20f

Increment

3.0 mm

0.7 mm

Image order

cr-ca

* The mAs should be adjusted to the body weight

of the child.

* The conversion factor for an 8-week-old, and a scan

range of 100 mm was used.

** The conversion factor for a 7-year-old and a scan

range of 200 mm was used.

Body weight

kV

mAs

CTDI

w

Effective dose

(mGy)

(mSv)

< 15 kg

80

30

0.8

Male: 0.4*
Female: 0.6*

15 – 24 kg

80

50

1.3

25 – 34 kg

80

75

1.9

Male: 0.9**
Female: 1.4**

35 – 44 kg

80

110

2.9

45 – 54 kg

80

150

3.9