Wholebodyangio – Siemens SOMATOM SENSATION CARDIAC A60 User Manual

Page 62

Advertising
background image

62

WholeBodyAngio

This scan protocol can be used for a whole Body Angio.

A range of 100 cm can be covered in 21.4 s.

WholeBodyAngio 2

nd

Reconstr.

kV

120

Effective mAs

130

Slice collimation

1.5 mm

Slice width

6 mm

2 mm

Feed/Rotation

24.0 mm

Rotation time

0.5 sec.

Kernel

B30f

B20f

Increment

6 mm

1.5 mm

Image order

cr-ca

The reconstruction can be loaded in the 3D Card,
MPR. Select the Range function and a predefined range,
linked to the series, will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.

MPRthick: WholeBodyAngio

coronal

Image thickness

3

Distance between images

3

Number of images

60

Advertising