Head – Siemens SOMATOM SENSATION CARDIAC A60 User Manual

Page 95

Advertising
background image

95

Head

ThinSliceSeq

RoutineSeq

kV

120

120

Effective mAs

250

250

Slice collimation 0.75 mm

1.5 mm

Slice width

4.5 mm

9.0 mm

Feed/Scan

9.0 mm

18.0 mm

Rotation time

0.5 sec.

0.5 sec.

Kernel

H31f

H31f

Image order

ca-cr

ca-cr

CTDI

w

58.2 mGy

60.3 mGy

Effective dose

Male: 0.7 mSv

Male: 1.6 mSv

Female: 0.8 mSvFemale: 1.8 mSv

PosteriorFossaOptimization Filter
PFO: To reduce beam-hardening artifacts in head
images, particularly in the base of the skull, use the
Posterior Fossa Optimization (PFO) filter.

If you want to set up the scan with only one range,
i. e. either 12 x 0.75 mm or 12 x 1.5 mm, you can
easily delete one mode by clicking the chronicle with
the right mouse button, and select cut.

Contrast medium IV injection
Start delay

60 sec.

Flow rate

2 ml/sec.

Total amount

50 – 60 ml