Heartview ct cascorespistd – Siemens SOMATOM SENSATION CARDIAC A60 User Manual

Page 32

Advertising
background image

32

HeartView CT

CaScoreSpiStd

Indications:

This is a standard spiral scanning protocol, using
an ECG gating technique for coronary calcium scoring
studies, with a rotation time of 0.42 seconds.

Topogram:
AP, 512 mm.
From the carina until
the apex of the heart.
A typical range of
15 cm covering the
entire heart can
be done in 11.6 s.

CaScoreSpiStd

kV

120

Effective mAs

133

Slice collimation

1.5 mm

Slice width

3 mm

Feed/Rotation

5.7 mm

Rotation time

0.42 sec.

Temporal Resolution

Up to105 ms*

Kernel

B35f

Increment

1.5 mm

Image order

cr-ca

CTDI

w

9.6 mGy

Effective Dose

Male: 2.1 mSv
Female: 3.1 mSv

* Depends on heart rate.

Advertising