HeadAngio

Indications:

Spiral mode for head CT angiography, e. g. cerebral
vascular abnormalities, tumors etc.

HeadAngio

2

nd

Reconstr.

kV

80

Effective mAs

*

Slice collimation

0.75 mm

Slice width

2.0 mm

1.0 mm

Feed/Rotation

6.8 mm

Rotation time

0.5 sec.

Kernel

H20f

H10f

Increment

2.0 mm

0.7 mm

Image order

ca-cr

* The mAs should be adjusted to the age of the child.

* The conversion factor for an 8-week-old, and a scan

range of 40 mm was used.

** The conversion factor for a 7-year-old and a scan

range of 60 mm was used.

Age

kV

mAs

CTDI

w

Effective dose

(mGy)

(mSv)

< 6 years

80

100

8.4

Male: 0.4*
Female: 0.4*

6 years –

80

150

12.6

Male: 0.3**

12 years

Female: 0.4**