Figure 3-8. inside region analog triggering mode, Figure 3-7, Above-high-level analog triggering mode -8 – National Instruments PCI-6110E/6111E User Manual

Page 28: Figure 3-8, Inside-region analog triggering mode -8

Advertising
background image

Chapter 3

Hardware Overview

PCI-6110E/6111E User Manual

3-8

© National Instruments Corporation

In above-high-level analog triggering mode, the trigger is generated
when the signal value is greater than highValue, as shown in Figure 3-7.
LowValue is unused.

Figure 3-7. Above-High-Level Analog Triggering Mode

In inside-region analog triggering mode, the trigger is generated when
the signal value is between the lowValue and the highValue, as shown
in Figure 3-8.

Figure 3-8. Inside-Region Analog Triggering Mode

highValue

Trigger

highValue

Trigger

lowValue

PCI_E.book Page 8 Thursday, June 25, 1998 12:55 PM

Advertising