Chapter 4

Signal Connections

© National Instruments Corporation

4-11

PCI-6110E/6111E User Manual

Differential Connections for Ground-Referenced
Signal Sources

Figure 4-3 shows how to connect a ground-referenced signal source to
a channel on the 611X E board.

Figure 4-3. Differential Input Connections for Ground-Referenced Signals

With this type of connection, the PGIA rejects both the common-mode
noise in the signal and the ground potential difference between the
signal source and the 611X E board ground, shown as V

cm

in Figure 4-3.

Differential Connections for Nonreferenced or
Floating Signal Sources

Figure 4-4 shows how to connect a floating signal source to a channel
on the 611X E board.

Ground-

Referenced

Signal

Source

Common-

Mode

Noise and

Ground

Potential

Instrumentation

Amplifier

V

m

Measured

Voltage

ACH0GND

-

+

-

+

V

s

V

cm

+

-

-

+

I/O Connector

ACH0+

ACH0-

ACH0 Connections Shown

PGIA

PCI_E.book Page 11 Thursday, June 25, 1998 12:55 PM