Figure 4-16. startscan input signal timing, Startscan signal -22, Figure 4-16. startscan input signal timing -22 – National Instruments PCI-6110E/6111E User Manual

Page 55: Startscan signal

Advertising
background image

Chapter 4

Signal Connections

PCI-6110E/6111E User Manual

4-22

© National Instruments Corporation

STARTSCAN Signal

Any PFI pin can externally input the STARTSCAN signal, which is
available as an output on the PFI7/STARTSCAN pin. Refer to
Figures 4-8 and 4-9 for the relationship of STARTSCAN to the DAQ
sequence.

As an input, the STARTSCAN signal is configured in the
edge-detection mode. You can select any PFI pin as the source for
STARTSCAN and configure the polarity selection for either rising or
falling edge. The selected edge of the STARTSCAN signal initiates a
scan. The sample interval counter starts if you select internally
triggered CONVERT*.

As an output, the STARTSCAN signal reflects the actual start pulse that
initiates a scan. This is true even if the starts are being externally
triggered by another PFI. You have two output options. The first is an
active high pulse with a pulse width of 25 to 50 ns, which indicates the
start of the scan. The second action is an active high pulse that
terminates at the start of the last conversion in the scan, which indicates
a scan in progress. STARTSCAN will be deasserted t

off

after the last

conversion in the scan is initiated. This output is set to tri-state at
startup.

Figures 4-16 and 4-17 show the input and output timing requirements
for the STARTSCAN signal.

Figure 4-16. STARTSCAN Input Signal Timing

Rising-edge

polarity

Falling-edge

polarity

t

w

t

w

= 10 ns minimum

PCI_E.book Page 22 Thursday, June 25, 1998 12:55 PM

Advertising