Figure 4-4. Differential Input Connections for Nonreferenced Signals

Figure 4-4 shows a bias resistor connected between ACH0 – and the
floating signal source ground. If you do not use the resistor and the
source is truly floating, the source is not likely to remain within the
common-mode signal range of the PGIA, and the PGIA will saturate,
causing erroneous readings. You must reference the source to the
respective channel ground.

Common-Mode Signal Rejection Considerations

Figure 4-3 shows connections for signal sources that are already
referenced to some ground point with respect to the 611X E board. In
this case, the PGIA can reject any voltage caused by ground potential
differences between the signal source and the board. In addition, with
differential input connections, the PGIA can reject common-mode noise
pickup in the leads connecting the signal sources to the board. The
PGIA can reject common-mode signals as long as V+

in

and V-

in

(input

signals) are both within ±11 V of the channel ground, for gain

1. For

gain <1, the input signals, for ACHO +, can be within ±42 V of the
channel ground.

+

-

+

Floating

Signal

Source

Instrumentation

Amplifier

V

m

Measured

Voltage

-

V

S

-

+

I/O Connector

ACH0GND

Bias

Current

Return

Paths

ACH0-

ACH0+

ACH0 Connections Shown

PGIA

1M

100pf

10nf

Bias
Resistor
(see text)