Figure 4-23. update* input signal timing, Figure 4-24. update* output signal timing, Uisource signal -28 – National Instruments PCI-6110E/6111E User Manual

Page 61: Uisource signal

Advertising
background image

Chapter 4

Signal Connections

PCI-6110E/6111E User Manual

4-28

© National Instruments Corporation

Figures 4-23 and 4-24

show the input and output timing requirements

for the UPDATE* signal.

Figure 4-23. UPDATE* Input Signal Timing

Figure 4-24. UPDATE* Output Signal Timing

The DACs are updated within 100 ns of the leading edge. Separate the
UPDATE* pulses with enough time that new data can be written to the
DAC latches.

The UI counter for the 611X E board normally generates the UPDATE*
signal unless you select some external source. The UI counter is started
by the WFTRIG signal and can be stopped by software or the internal
Buffer Counter.

D/A conversions generated by either an internal or external UPDATE*
signal do not occur when gated by the software command register gate.

UISOURCE Signal

Any PFI pin can externally input the UISOURCE signal, which is not
available as an output on the I/O connector. The UI counter uses the
UISOURCE signal as a clock to time the generation of the UPDATE*
signal. You must configure the PFI pin you select as the source for the
UISOURCE signal in the level-detection mode. You can configure the

Rising-edge

polarity

Falling-edge

polarity

t

w

t

w

= 10 ns minimum

t

w

t

w

= 50-75 ns

PCI_E.book Page 28 Thursday, June 25, 1998 12:55 PM