Chapter 2

Installation and Configuration

© National Instruments Corporation

2-5

NI PXIe-1065 User Manual

rated ambient temperature as stated in Appendix A,

Specifications

. If the

temperature exceeds the stated spec the power switch LED will blink green,
as discussed in the

Power Inhibit Switch LED Indicator

section of this

chapter.

Setting Fan Speed

The fan-speed selector switch is on the rear panel of the NI PXIe-1065
chassis. Refer to Figure 1-2,

Rear View of the NI PXIe-1065 Chassis

, to

locate the fan-speed selector switch. Select High for maximum cooling
performance or Auto for improved acoustic performance. When set to
Auto, the fan speed is determined by chassis intake air temperature.

Installing Filler Panels

To maintain proper module cooling performance, install filler panels
(provided with the chassis) in unused or empty slots. Secure with the
captive mounting screws provided.

Installing Slot Blockers

The cooling performance of the chassis can be improved by installing
optional slot blockers. Refer to

ni.com

for more details.

Rack Mounting

Rack mount applications require the optional rack mount kits available
from National Instruments. Refer to the instructions supplied with the rack
mount kits to install your NI PXIe-1065 chassis in an instrument rack.
Refer to Figure A-3,

NI Chassis Rack Mount Kit Components

.

Note

You may want to remove the feet from the NI PXIe-1065 chassis when rack

mounting. To do so, remove the screws holding the feet in place.

Connecting Safety Ground

Caution

The NI PXIe-1065 chassis are designed with a three-position NEMA 5-15 style

plug for the U.S. that connects the ground line to the chassis ground. To minimize shock
hazard, make sure the electrical power outlet you use to power the chassis has an
appropriate earth safety ground.