background image

Operation

97

4

Operation

4.1

Switch on the instrument

Note
The instrument has an energy saving feature to avoid
unnecessary battery depletion.
The energy saving feature switches the instrument off if no
key has been pressed for an hour.
The energy saving feature is not active:

l

if the power is supplied by the plug-in power supply

l

if the

AutoStore

function is active

l

if a PC is connected

l

if the recorder cable is connected

l

if the printer cable is connected
(for external printers)

1 Place the instrument on a flat surface and protect it

against intense light and heat.

2 Connect the conductivity measuring cell to the instru-

ment.

3 Press the <on/off> key.

The

display test

appears briefly on the display.

The instrument then switches automatically to the
previously selected measuring mode.

4 Check or determine the cell constant [C]. The proce-

dure is described in section 4.3, on page 106.