Operation

129

Setting the

data transmission

interval

The interval is set to OFF in the factory.
To start the data transmission, set up an interval (5 s, 10 s,
30 s, 1 min, 5 min, 10 min, 15 min, 30 min, 60 min):

Note
If the

AutoStore

function is active, the data transmission is

performed according to the setting of the storage interval
(Int1). Set the storage interval (Int1) to OFF to activate the

data transmission

interval (Int2).

Note
You can also set the

data transmission

interval (Int2) in the

Configuration

menu (see page 128).

1 Press and hold down the <run/enter> key.

2 Press the <auto read> key.

Int 2 appears on the display.

3 Press <

> <

> to set up the required time interval

between storage processes.

4 Confirm with <run/enter>.

The instrument changes automatically to the measur-
ing mode.