Operation

132

4.6

Configuration

You can adapt the conductivity meter to your individual
requirements. To do this, the following parameters can be
called up/changed (the status on delivery is marked in bold):

Note
You can leave the configuration menu at any time.
Parameters that have already been changed are stored.
To do this, press the <

ì

> key.

Baud rate

1200, 2400, 4800. 9600

Data transmission inter-
val (Int 2)

OFF, 5 s, 10 s, 30 s,
1 min, 5 min, 10 min,
15 min, 30 min, 60 min

Calibration interval (Int 3) 1 ... 180 ... 999 d

AutoRange ARng

yes, no

TREF 25/TREF 20

t25 or t20

Measured values given
as conductivity or resi-
stance values

S/cm or M

Temperature unit

°C or °F

Date/time

as required

1 Switch off the instrument.

2 Press and hold down the <

ì

> key.

3 Press the <on/off> key.

The display test

appears briefly on the display.

The instrument then switches automatically to the
setting of the baud rate.