Xylem CG 853P User Manual

Page 37

Advertising
background image

Operation

113

Setting up the

temperature

compensation TC

The calculation of the temperature compensation is based
on the preset reference temperature, Tref 20 or Tref 25 (see
chapter 4.6 C

ONFIGURATION

).

You can select one of the following temperature compensa-
tions:

l

Non-linear temperature compensation "nLF"
according to DIN 38404 or EN 27 888

l

linear temperature compensation "Lin" with a
coefficient that can be set in the range 0.001 ... 3.000 %/K

l

no temperature compensation

Note

ips

Select the following temperature compensations to work
with the test samples given in the table:

Application

notes

Test sample

Temperature
compensation TC

Display
indicator

Natural water
(ground water,
surface water,
drinking water)

nLF
according to DIN 38404
EN 27 888

Ultrapure water

nLF
according to DIN 38404
EN 27 888

Other aqueous
solutions

Set linear
temperature coefficient
0.001 ... 3.000 %/K

Salinity
(seawater)

Automatically nLF accord-
ing to IOT

Sal,