Xylem CG 853P User Manual

Page 32

Advertising
background image

Operation

108

Note
This method of automatically determining the cell constant
by calibrating in the 0.01 mol/l KCL control standard can
only be used for measuring cells with a cell constant in the
range 0.450 ... 0.500 cm

-1

or 0.800 ... 1.200 cm

-1

.

Calibration

evaluation

After calibrating, the instrument automatically evaluates the
current condition of the calibration.
The evaluation appears on the display.

Calibration protocol

The calibration protocol contains the calibration data of the
current calibration.

Note
You can automatically print out a calibration protocol after
the calibration. To do so, switch on the printer before the cal-
ibration (Print LED lights up green). After a valid calibration,
the protocol is printed.

Display

Cell constant [cm

-1

]

0.450 ... 0.500 cm

-1

0.800 ... 1.200 cm

-1



Perform error elimination
according to chapter
6 W

HAT

TO

DO

IF

...

outside of the ranges
0.450 ... 0.500 cm

-1

or
0.800 ... 1.200 cm

-1

CALIBRATION PROTOCOL
14.04.99 11:37
Device No.: 99990000
CALIBRATION CONDUCTIVITY
Cal Time: 14.04.99 / 11:37
Cal Interval: 180d
Cal Std.: 0.01 mol/l KCL
40.0 °C
Conduct./Tref25: 1413µS/cm
Cell Const : 0.975 1/cm
Probe : +++