Xylem CG 853P User Manual
Page 49

Operation
125
The printout
contains:
l
Calibration protocol:
– Date/time of the printout
– Instrument number (Device No.)
– Date/time of the calibration (Cal Time)
– Calibration interval (Cal Interval)
– Calibration standard 0.01 mol/l KCL (Cal Std.) [°C/F]
– Conductivity of the calibration standard at 25 °C
(Conduct./Tref25)
– Cell constant (Cell Const)
– Probe evaluation (Probe)
Sample printout
CALIBRATION PROTOCOL
14.04.99 11:37
Device No.: 99990000
CALIBRATION CONDUCTIVITY
Cal Time: 14.04.99 / 11:37
Cal Interval: 180d
Cal Std.: 0.01 mol/l KCL
23.0 °C
Conduct./Tref25: 1413 µS/cm
Cell Const : 0.975 1/cm
Probe: +++
No. 1:
01.01.99 00:04
2.40 mS/cm 25 °C
Tman
nLF
Tref25 C = 0.475 1/cm
Ident : 1
No. 2:
10.01.99 10:09
2.40 mS/cm 25.3 °C
Tauto
nLF
Tref25 C = 0.475 1/cm
Ident : 1
No. 3:
12.01.99 01:48
2.40 mS/cm 21.6 °C
Tauto
nLF
Tref25 C = 0.475 1/cm
Ident : 1